e.cam (Single / Dual Head)
Technical Specifications |
| Minimum Room Dimensions |
|
| Minimum Ceiling Clearance |
|
| Gantry Weight (w/o Collimators) |
|
| Room Floor Loading (w/o Collimators) |
|
| Gantry Floor Loading "Moving Component" Worst Case w/Heaviest Collimator(s) |
150 lbs/ft² |
| Crystal Thickness |
|
| Table Weight Limit |
400 lbs |
| Energy Windows |
|
| Collimator Exchange |
Semi-Automatic |
| Detector Separation |
|
| Upright Imaging Height |
|
| Temperature / Humidity Requirements |
60°-95° F, 15-80% (Non-Condensing) |
| THVAC Requirements |
Less Than ±8° F Variation per Hour |
| Heat Generation |
6,800 BTU/hr |
| Voltage / Current |
200-240V 15A |
| Peak / Quiescent Current (est.) |
13.4A, 10A |
| Floor Level Specification |
|